Crystal EX



Adrona Crystal EX produces ultrapure and pure water. This multipurpose water purification system is highly appreciated due to the very affordable price.

Ordering information :
Model Part number
Crystal EX Trace EX-1001
Crystal EX HPLC EX-1101
Crystal EX Bio EX-1201
Consumables
Part number Description Replacement criteria Comments
10320 Replacement pre-filter If the filters are clogged or every 6 months
10310 Replacement deionization module Grade 2 water conductivity is > 0.5 μm/cm constantly or every 12 months
10030 Polishing module “Polishing+” Grade 1 water conductivity is > 0.1 μm/cm constantly or every 12 months
10017 Replacement sterilization UV bulb When required (on average every 2 years) „Bio” systems only
10018 Replacement photooxidation UV bulb 2 years on average „HPLC” and „Bio” systems only
10013 Replacement 0.22 μm dispense microfilter Every 6–12 months „Trace” and „HPLC” systems
10120 Replacement ultrafilter Every 3–6 months „Bio” systems only

Description EX series
  Trace HPLC Bio
Water type
  • ultrapure water (Grade 1)
  • pure water (Grade 2)
  • ultrapure water (Grade 1)
  • pure water (Grade 2)
  • ultrapure water (Grade 1)
  • pure water (Grade 2)
Application
  • atomic absorption spectrometry
  • plasma optical emission spectrometry
  • other inorganic trace analysis
  • chromatography
  • mass spectrometry
  • microbiology
  • molecular biology
  • highly sensitive biology applications
Display
Monochrome LCD display
Conductivity sensor
*
*
*
TOC Monitor
-
-
-
Volumetric dispensing
-
-
-
Connection to Flow point
-
-
-
Storage tank
required but not included
Installation
installable on a laboratory bench
Flow diagram
Specifications
Crystal EX Trace
Crystal EX HPLC
Crystal EX Bio
Grade 1 water resistivity at 25 °C
18.2 MΩ x cm
18.2 MΩ x cm
18.2 MΩ x cm
Grade 1 water conductivity at 25 °C
0.055 μS/cm
0.055 μS/cm
0.055 μS/cm
Grade 2 water resistivity at 25 °C
>10 MΩ x cm
>10 MΩ x cm
>10 MΩ x cm
Grade 2 water conductivity at 25 °C
<0.1 μS/cm
<0.1 μS/cm
<0.1 μS/cm
Total Organic Carbon (TOC) level
<10 ppb
<5 ppb*
<5 ppb*
RNase
-
-
<0.01 ng/mL
DNase
-
-
<4 pg/μL
Bacteria
<0.01 CFU/mL
<0.01 CFU/mL
<0.01 CFU/mL
Endotoxins
<0.15 EU/mL
<0.15 EU/mL
<0.001 EU/mL
Particles >0.22 μm
<1 per mL
<1 per mL
<1 per mL
Nominal flow, pure water (to storage tank)
10 L/h
10 L/h
10 L/h
Dimensions (WxDxH), cm
40x35x55
40x35x55
40x35x55
System weight, kg
17
18
19
Operation weight, kg
24
25
26
Feed water pressure
0.4 – 4 bar
0.4 – 4 bar
0.4 – 4 bar
Feed water conductivity
< 1500 μS/cm
< 1500 μS/cm
< 1500 μS/cm